PSP103型氣相色譜儀 執(zhí)行標準:GB/T 17623-1998 GB/T 7252-2001 DL/T 722-2000 PSP103型氣相色譜儀是本公司研制的新一代氣相色譜儀。儀器采用微機自動控制、全鍵盤操作、大屏幕液晶中文顯示,具有控溫精度高、性能穩(wěn)定可靠、靈敏度高、重復(fù)性好等特點。 執(zhí)行標準:GB/T 17623-1998《絕緣油中溶解氣體組分含量的氣相色譜測定法》 GB/T 7252-2001《變壓器油中溶解氣體分析和判斷導(dǎo)則》 DL/T 722-2000《變壓器油中溶解氣體分析和判斷導(dǎo)則》 儀器采用雙柱并聯(lián)分流系統(tǒng)。配有TCD檢測器和雙FID檢測器以及甲烷轉(zhuǎn)化器,能一次進樣實現(xiàn)油中溶解氣體九組分(H2、O2、N2、CH4、C2H2、C2H4、C2H6、CO、CO2)的全分析??捎糜陔娏ο到y(tǒng)油氣設(shè)備內(nèi)部故障檢測,氫冷發(fā)電機冷卻介質(zhì)分析、六氟化硫(SF6)雜質(zhì)分析、鍋爐煙氣分析、天然氣分析和環(huán)境監(jiān)測分析等。既可作分析,又可作通用分析,因而成為石油、化工、礦山等系統(tǒng)作氣體分析的zui有效設(shè)備。
1、溫度范圍 柱室:室溫加10℃~399℃ 熱導(dǎo)檢測器:室溫加10℃~300℃ 氫焰檢測器:室溫加10℃~350℃ 轉(zhuǎn)化爐:室溫加10℃~399℃ 溫度控制精度:±0.1℃ 溫度顯示精度:0.1℃ 2、TCD檢測器 TCD靈敏度:≥5000mv.ml/mg(正十六烷/異辛烷) 噪聲:≤0.02mV 漂移:≤0.05mV/30min 3、FID檢測器 檢測限:≤5×10-12g/s(正十六烷/異辛烷) 噪聲:≤0.02mV 漂移:≤0.05mV/30min 4、定性重復(fù)性誤差:≤1% 5、定量重復(fù)性誤差:≤3% 6、zui低檢測量 H2 :≤2ppm CO、CO2:≤2ppm 烴類:≤0.1 ppm
The PSP103 Gas chromatograph Implementation of the standards: GB / T 17623-1998 GB / T 7252-2001 DL / T 722-2000 PSP103 type gas chromatograph is the latest development of a new generation of gas chromatograph. The instrument uses a microcomputer automatic control, full keyboard, large screen LCD Chinese display with high precision temperature control, stable and reliable performance, high sensitivity, good repeatability. Implementation of the standards: GB / T 17623-1998 "component content of dissolved gases in the insulating oil gas chromatographic method GB / T 7252-2001 "transformer oil dissolved gas analysis and judgment guide DL / T 722-2000 "transformer oil dissolved gas analysis and judgment Guidelines The instrument uses a double column parallel shunt system. Equipped with a TCD detector and dual FID detector, as well as the conversion of methane, a single injection Dissolved Gas nine components (H2, O2, N2, CH4, C2H2, C2H4, C2H6, CO, CO2) of the full analysis. Power system can be used for oil and gas equipment internal fault detection, hydrogen-cooled generator cooling medium, and sulfur hexafluoride (SF6) impurity analysis, boiler flue gas analysis, gas analysis and environmental monitoring and analysis. Both exclusive analysis, but also for the general analysis, and thus become the most effective equipment for the petroleum, chemical, mining and other systems for gas analysis. 1, the temperature range The column room: room temperature and 10 ℃ ~ 399 ℃ Thermal conductivity detector: room temperature and 10 ℃ ~ 300 ℃ Hydrogen flame detector: room temperature plus 10 ° C to 350 ° C The conversion furnace: room temperature plus 10 ℃ ~ 399 ℃ Temperature control accuracy: ± 0.1 ℃ Temperature display accuracy: 0.1 ° C 2, TCD detector TCD sensitivity: ≥ 5000mv.ml/mg (n-hexadecane / isooctane) Noise: ≤ 0.02mV Drift: ≤ 0.05mV/30min 3, FID detector Detection limit: ≤ 5 × 10-12g / s (n-hexadecane / isooctane) Noise: ≤ 0.02mV Drift: ≤ 0.05mV/30min Qualitative repeatability error: ≤ 1% Quantitative repeatability error: ≤ 3% 6, the lowest detectable amount H2: ≤ 2ppm CO, CO2: ≤ 2ppm Hydrocarbons: ≤ 0.1 ppm |